Macherey-Nagel Photometers

Nanocolor® Vis Spectrophotometer

  • Automatic cuvette detection & recognition
  • Automatic method & wavelength selection
  • Auto measurement & data storage
  • No cuvette slot cover needed
  • Universal cuvette slot - no adaptors needed
  • Self explanatory user guidance displayed
  • Easy use backlit touch screen
  • Manual with pictogram instruction
  • Documentation of results according to GLP
  • Clear Memory management
  • Optional sample allocation with hand held scanner
  • Internal QA to Iso 9001
  • >200 preprogrammed tests
  • Up to 100 user defined tests possible
  • SAC at 436nm
  • Colour determination DIN EN ISO 7887
  • Nephelometric turbidity EN ISO 7027
  • Full Spectrum analysis

The NANOCOLOR® VIS from MACHEREY-NAGEL is a powerful VIS spectrophotometer with monochromator (340-1100 nm) for universal use in all areas of water and waste water analysis. This includes municipal and industrial waste water, drinking water, process water, surface water, ground water as well as cooling and boiler feed water. The NANOCOLOR® VIS is also the ideal test unit for quality control in various fields of industry, such as food and beverage industries.

Select a test type:

Or Return to

Macherey-Nagel Index page

 
NANOCOLOR Vis Spectrophotometer Features

No cuvette adapter required

Benefits:

  • Reduced analysis time/increased throughput
  • Increased Accuracy
  • Secure assured results
  • Flexibility of use

Applications:

  • Municipal & Industrial wastewater
  • Potable water
  • Process water
  • Surface water
  • Ground water
  • Boiler feed water
  • Cooling water

Technical data

  • Type: VIS Spectrophotometer with reference detector technology (RDT)
  • Light sources: Halogen lamp
  • Optical system: Monochromator
  • Wavelength range: 340-1100 nm
  • Wavelength accuracy: ± 1 nm
  • Wavelength resolution: 0.3 nm
  • Wavelength calibration: Automatic
  • Wavelength selection: Automatic, barcode, manual
  • Scan speed: 1 complete scan in less than 1 min
  • Spectral bandwidth: < 4 nm
  • Photometric range: ± 3.0 E in wavelength range 400-900 nm
  • Photometric accuracy: 0.005 E at 0.0 - 0.5 E; 1% at 0.5 - 2.0 E
  • Photometric linearity: < 0.5% at 2 E; = 1% at > 2 E
  • Stray light: < 0.5%
  • Measuring modes: More than 200 pre-programmed tests, 100 optionally programmable methods, absorbance, transmission, factor, kinetics, 2-point calibration, scan, nephelometric turbidity measurement
  • Cuvette holder: Test tubes 16 mm OD, rectangular cuvettes 2, 10, 20, 50 mm
  • Data memory: 1000 measured data sets, GLP conform
  • Display: Coloured, backlit LCD touchscreen
  • Operation: Barcode technology, display user guidance, touchscreen
  • Languages: de, en, fr, es, nl, it, hu, pl, pt, cz
  • External light: Insensitive, open cuvette slot
  • Interfaces: USB and bi-directional serial RS 232
  • Update: Via Internet / PC and USB stick (included in delivery)
  • Operating range: 10 - 40 °C , max. 80% relative humidity (without condensation)
  • Power supply: 12 V DC / 2.0 A
  • Dimensions: L / W / H: 325 / 215 / 115 mmWeight: 2 kg
  • Warranty: 2 years

Universal cuvette slot...

...for the use of round tubes (16 mm OD)...

...and rectangular cuvettes (2, 10, 20, 50 mm).

Envitech Products

MJK products


The Envitech Sampsys Autosampler