Type:
VIS Spectrophotometer with reference detector
technology (RDT)
-
Light
sources: Halogen lamp
-
Optical
system: Monochromator
-
Wavelength
range: 340-1100 nm
-
Wavelength
accuracy: ± 1 nm
-
Wavelength
resolution: 0.3 nm
-
Wavelength
calibration: Automatic
-
Wavelength
selection: Automatic, barcode, manual
-
Scan
speed: 1 complete scan in less than 1 min
-
Spectral
bandwidth: < 4 nm
-
Photometric
range: ± 3.0 E in wavelength range
400-900 nm
-
Photometric
accuracy: 0.005 E at 0.0 - 0.5 E; 1% at
0.5 - 2.0 E
-
Photometric
linearity: < 0.5% at 2 E; = 1% at >
2 E
-
Stray
light: < 0.5%
-
Measuring
modes: More than 200 pre-programmed tests,
100 optionally programmable methods, absorbance,
transmission, factor, kinetics, 2-point calibration,
scan, nephelometric turbidity measurement
-
Cuvette
holder: Test tubes 16 mm OD, rectangular
cuvettes 2, 10, 20, 50 mm
-
Data
memory: 1000 measured data sets, GLP conform
-
Display:
Coloured, backlit LCD touchscreen
-
Operation:
Barcode technology, display user guidance, touchscreen
-
Languages:
de, en, fr, es, nl, it, hu, pl, pt, cz
-
External
light: Insensitive, open cuvette slot
-
Interfaces:
USB and bi-directional serial RS 232
-
Update:
Via Internet / PC and USB stick (included in delivery)
-
Operating
range: 10 - 40 °C , max. 80% relative
humidity (without condensation)
-
Power
supply: 12 V DC / 2.0 A
-
Dimensions:
L / W / H: 325 / 215 / 115 mmWeight: 2 kg
-
Warranty:
2 years