The Isco EZ
TOC uses low-temperature UV-promoted, persulfate oxidation to convert
organic carbon compounds to CO2, which is then measured using a
non-dispersive infrared detector to determine the total organic
carbon. Sample is drawn into the EZ TOC by the source pump, and
continuously sampled by a feed pump to a scrubbing column, where
dilute phosphoric acid converts inorganic carbon compounds to CO2
which is removed by sparging.
The
scrubbed sample is pumped to the UV reactor where organic carbon
compounds are oxidized to CO2 by interaction of the sample with
the UV light, carrier gas, and concentrated persulfate. The CO2
is dried and measured by a non-dispersive infrared detector (NDIR),
and the value converted to TOC in mg/L.
Up to eight
programmable relays can provide a remote indication of system operation
and report trouble conditions. A fail-safe shutdown feature prevents
damage in case of a malfunction. Manual inhibit buttons let you
disable alarms, hold an analog output at the current reading, or
bypass scheduled cleaning and calibrations.
In addition
to being a continuous on-line process analyzer, the EZ TOC allows
you to conveniently measure grab samples using the Manual SampleValve.
Standard
Features
Analysis
method complies with USEPA, ASTM, ISO and CEN standards for TOC
determination.
Flexible
four-motor pump system. Advanced hydraulics allow for unlimited
number of ranges between 10 mg/l and 10,000 mg/l TOC.
Four separate,
programmable form Type C relays can be activated on up to 18 parameters
monitored, including two alarm levels, loss of sample or gas flow,
leak detection and more.
User-selectable
and scalable 4-20 mA or 0-20 mA output.
RS-232 digital
output.
Large, backlit
display makes it easy to check operating status in low light conditions
without opening the door.
Built-in
reagent timer alerts the operator of upcoming chemical requirements.
Programmable,
automatic cleaning and calibration cycles provide accurate, low-maintenance
operation.
High efficiency
single-lamp UV reactor reduces costs and provides positive indication
of lamp failure, instead of producing inaccurate data as occurs
when a lamp fails in a multi-lamp system.
Enhanced
NDIR detector uses the same infrared sources and detectors for
the sample and reference cells, providing a stable, low-noise
CO2 detection.
Applications
Increase
the efficiency of water/wastewater treatment to reduce process
costs in both industrial and municipal applications.
Provide warning
of organic spills, to protect expensive processes and equipment
in addition to monitoring for regulatory compliance.
List of
articles - On-line TOC applications using EZtoc analyser
Options
and Accessories
PA2 or PA3
Sample Filter - Required where sample contains particles larger
than 200 microns.
Isco Carrier
Gas Converter - cleans and reduces the CO2 content of compressed
air. This accessory must be installed in the air supply line when
on-site compressed air is used.
Extra relay
board. Adds four additional Type C relay outputs to the four standard
relays.
High-TIC
and ultra-high TIC removal kits for applications where TIC levels
exceed maximum expected TOC level.
Specifications
of EZtoc TOC analyser
Function:
Total Organic Carbon (TOC) or Total Carbon (TC)
Measurement
ranges: Five full-scale ranges between 10 and 10,000 mg/L carbon.
Specify.
Data handling:
EIA/TIA-232E and CCITT V.28 Serial (unidirectional )
Electrical
output: Programmable, user-selectable 4-20mA or 0-20mA current
loop into 1000 ohms maximum
Alarms: Two
programmable alarm levels driving up to 8 programmable Type C
relays (4 standard; 4 optional)
Programmable
outputs: Up to 8 customer programmable outputs on Type C relays
(4 relays are standard). Can be programmed to output any combination
of 18 system parameters (includes the two alarms). Each relay
output board rated at 0.5A @ 24VDC/240VAC.
Calibration:
2 point liquid, 2 point gas, manual and automatic
Performance
Characteristics
Response
time: Less than 8 minutes to T90 at 100 mg/L configuration TOC
Repeatability:
± 1% full scale
Drift: ±
1% of full scale over 72 hours without calibration at 20°C
Temperature
stability: Less than 2% of full scale drift over ambient range
of 10° C to 30°C without liquid calibration
Inorganic
carbon removal: > 95% removal when TIC equals full scale TOC
with standard TIC scrubber
Accuracy
(% of full scale): ±1.5% for TOC concentrations from 0%
to 75% of full scale. ± 2.5% for TOC concentrations from
75% to 100% of full scale
Sample Requirements
Inlet pressure:
Atmospheric to 3 psi (0 - 0.21Bar)
Drain pressure:
Gravity drain, vented to atmosphere
Suspended
solids: PA2 or PA3 Sample Preparation/Filter System required where
particle size exceeds 200 µm. 3% maximum concentration of
suspended solids by volume. Oxidation rate is dependent on constituency
Flow rate
of sample through unit: 35 ml/min. maximum at 60 Hz
Carrier Gas:
Compressed air containing < 3mg/L CO2.
Consumption:
Less than 750 cc/min. at 30 psi
Reagents:
1.5 M Na2S2O8 (sodium persulfate) & 10% v/v H3PO4 (phosphoric
acid) both in deionized water. Consumption is 5.2 gallons/month
each.
Physical
Characteristics
Enclosure:
NEMA 4, IP 65
Size (HxWxD):
30 x 24 x 22 inches (76 x 61 x 56 cm)
On-line
BOD for process control
"We
also review the analyzer data on a monthly basis and compare the results
with routine laboratory BOD5 data for the same time period. We have
found there to be an excellent correlation between the lab and the
analyzer results."