A
COMPARISON OF THE BOD5 and the continuous SHORT-TIME BOD (BOD-M3),
by M Kohne, Siegen; F W Siepmann, Reinheim and D te Heesen, Dinslaken
Abstract
:
The
BOD5 and the BOD-M3 were compared using a suitably chosen test
procedure. The investigation shows that clearly defined objectives
are critical in comparative investigations in the analysis of
waste water. A discontinuous load, in the form of a step curve
with two concentration levels and a predefined ratio of these
levels to each other, has proven useful in this investigation
and in a calibration procedure which has resulted from it.
The results of the comparative measurement of the BOD5, as a round-robin
experiment, were very scattered, as was expected. The respective
laboratory averages, with a deviation of +10 to 15% , are method-dependent
and are acceptable.
The measurement of the BOD-M3 in the experiment depicts the input
loading as a step curve. The required stable curves are maintained.
The predefined ratio of 3.0 is exceeded by 5%, the BOD-M3 is determined
essentially by variations of the mixing pumps and in the oxygen
measurements. The reproducibility for the detection of a biologically
degradable substance is between 95 and 98%. The accuracy in reproducing
the BOD-M3 cannot be achieved with the BOD5. It is therefore useful
to indicate not only the BOD-M3 calibrated using the BOD5, but
also the uncalibrated standard value of the BOD-M3.